Hydraulic universal testing machine is prone to problems and solutions

Hydraulic universal testing machine is prone to problems and solutions

1. Hydraulic universal testing machine stalls or does not move

1 After the photoelectric switch at the rear of the motor is worn for a long time, the receiving part of one side is worn.

Solution: replace the photoelectric switch

2 software parameter settings have problems

Solution: Reset the parameters of the software

2. Test machine force value zero point fluctuation (zero drift)

After several years of use of the test machine, the force value zero point will generally fluctuate up and down.

1 oil circuit blockage (generally, when there is no movement, there is zero fluctuation)

Solution: Re-replace oil, clean the oil circuit

2 The voltage used by the machine is unstable

Solution: use a voltage regulator to adjust the voltage

3 sensor has collision damage

Solution:

a, replace the sensor

b. Reset the test machine.

3, the test machine measurement results produce errors

1 The main part of the main unit is not installed horizontally, which will cause friction between the working piston and the working cylinder wall, resulting in errors. It generally exhibits a positive difference and gradually becomes smaller as the error of the load.

2 The dynamometer part of the dynamometer is not installed horizontally, which will cause friction between the pendulum shaft bearings, which is generally realized as a negative difference. The above two kinds of errors have a large influence on the small load measurement, and the impact on the large load measurement is relatively small.

Solution:

a. First check whether the test machine is installed horizontally, and level the horizontal direction of the main engine with the frame level on both sides of the working cylinder (or column).

b. Adjust the level of the dynamometer before and after the dynamometer on the front of the pendulum, and align the edge of the pendulum with the inner scribe line. Use the horizontal ruler to adjust the left and right horizontal of the body against the side of the oscillating bar.

Global hardware network

Concerned about surprises

Label: Hydraulic universal testing machine is prone to problems and solutions